Thermal conductivity measurement
Stationary plate method
- Stationary plate method based on DIN EN 12664
- Surface temperatures and heat flux densities are measured in a stationary state
- Determination of the effective thermal conductivity of homogeneous and inhomogeneous test specimens
- Measurement is carried out in a climatic chamber under defined conditions
Schematischer Aufbau Plattenmethode
| Technical data | |
|---|---|
| Samples | Solids plate-shaped with plane-parallel surfaces |
|
Sample dimensions |
Area depending on heat flow plates at least |
| Temperature range | -10 °C to 60 °C |
| Measuring accuracy | ±6 % for Styrodur reference (≈0.035 W m-1 K-1 ) |
Transient hot wire method
- Transient method with hot wire for heat supply and measurement of the temperature change
- Determination of thermal conductivity from analytical solution of the heat conduction equation assuming a line source in the infinitely extended body
Hitzdrahtsonde - (1) Hitzdraht, (2) Halterung, (3) Thermoelement, (4) elektr. Anschlüsse
| Technical data | |
|---|---|
| Samples | Powders, bulk solids, liquids, solids (after solidification around probe) |
|
Sample dimensions |
Cylindrical measuring container |
| Temperature range | -10 °C to 60 °C |
| Measuring accuracy | ±5 to ±10 % |
Applied Precision Isomet 2114 measuring device
- Transient method for measuring thermal conductivity and volumetric heat capacity
- Insertion probe for powders, bulk solids, solids and contact probe for solids
- Measurement carried out in a climatic chamber under defined conditions
Messgerät Applied Precision Isomet 2114
| Technical data | |
|---|---|
| Samples | Powder, bulk solids, solids |
|
Sample dimensions |
Insertion probe |
| Temperature range | -20 °C to 70 °C |
| Measuring accuracy | ±5 % in the range 0.015 to 0.70 W m-1 K-1 ±10 % in the range 0.7 to 6.0 W m-1 K-1 |